• Raising the bar in orthopedic and spinal instrumentation.
• Accuracy and repeatability under the most demanding surgical conditions.
• Unique drive mechanism eliminates reverse torque action upon reaching peak.
• Individually serialized upon customer request; for ease of traceability and recalibration cycles.
• ALL our Total-Control torque-limiting wrenches are guaranteed to be within +/- 5% of the customer specified torque setting.
![]() |
|||||||||||||||||
![]() |
|||||||||||||||||
![]() |
|||||||||||||||||
![]() |
|||||||||||||||||
** Patent Pending
Click here to review the detailed listing of Bradshaw Medical, Inc. patents and trademarks.
Recent testing confirms that our torque limiting drivers maintain their accuracy and repeatability. The graph to the right details test results for a T-Handle driver set to a nominal torque of 90 lb-in., then subjected to an extreme 3,000 clicks without rest periods, cleaning, or sterilization. Throughout the test, nominal torque remained within a tolerance of +/- 5% of the original pre-set torque, with an average nominal torque value of 91.577.

![]() |
Now Available! Silicone Overmolded Adapter Collars** |
|